ELECTRICAL CONNECTIONS
24
TIDALFLUX 2300 F
www.krohne.com
06/2013 - 4002490202 MA TIDALFLUX 2300 F R02 en
4.4 Signal cable B (type BTS 300), construction
• Signal cable B is a triple-shielded cable for signal transmission between the measuring
sensor and signal converter.
• Bending radius: ≥ 50 mm / 2"
4.5 Signal cable A (type DS 300), construction
• Signal cable A is a double-shielded cable for signal transmission between the measuring
sensor and signal converter.
• Bending radius: ≥ 50 mm / 2"
Figure 4-11: Construction of signal cable B
1 Stranded drain wire for the inner shield (10), 1.0 mm
2
Cu / AWG 17 (not insulated, bare)
2 Insulated wire (2), 0.5 mm
2
Cu / AWG 20 with stranded drain wire (20) of shield
3 Insulated wire (3), 0.5 mm
2
Cu / AWG 20 with stranded drain wire (30) of shield
4 Outer sheath
5 Insulation layers
6 Stranded drain wire (6) for the outer shield (60), 0.5 mm
2
Cu / AWG 20 (not insulated, bare)
Figure 4-12: Construction of signal cable A
1 Stranded drain wire (1) for the inner shield (10), 1.0 mm
2
Cu / AWG 17 (not insulated, bare)
2 Insulated wire (2), 0.5 mm
2
Cu / AWG 20
3 Insulated wire (3), 0.5 mm
2
Cu / AWG 20
4 Outer sheath
5 Insulation layers
6 Stranded drain wire (6) for the outer shield (60)