beautypg.com

0 appendix a – communication device, 1 install and open the software, 2 connect the probe to the communication device – In-Situ RDO PRO Process Optical Dissolved Oxygen Sensor Manual User Manual

Page 25: Appendix a – communication device, Install and open the software, Connect the probe to the communication device

background image


10.0 Appendix A – Communication Device

The Communication Device is an accessory product that can be used to calibrate and set up RDO
probes.

10.1 Install and Open the Software

The Comm Kit Software must be installed on a computer before you connect to the probe.

10.2 Connect the Probe to the Communication Device

The Communication Device connects a stripped-and-tinned probe to a computer via USB
connection.

1. Disconnect the instrument from the PLC.

2. The communication device includes an electrical connection diagram label. To attach the

instrument to the communication device, depress a lever and insert the appropriate wire in the
location specified by the diagram.

3. Attach the USB connector to a USB port on the computer. Follow the directions provided in the

Communication Device Kit to set up the probe.

25

Appendix A – Communication Device

This manual is related to the following products: