Page 22

background image


12. Write the Calibration Mode Off command (0xE002) to the sensor command register to place the

sensor in normal operation. If the calibration mode is turned off without a calibration update
command, or the calibration command returned an exception, the previous calibration shall be
restored.

13. Optional: If you saved the temperature and saturation parameter units at the start of the

process, write the original values back.

14. Optional: If you saved the Sensor Data Cache Timeout register 9463 at the start of the process,

write the original value back.

6.5.1

Calibration Calculations

Calibrated oxygen reading:

O

2RC

= c

o

+ c

1

× O

2RU

Where:

c1 = (O

2

100%Sat)/(

O

2RUS

− O

2RUZ

)

c

o

= -c

1

× O

2RUZ

Where:

O

2

100%Sat is the theoretical 100% saturation point

O

2RUS

is the un-calibrated reading at 100% saturation

O

2RUZ

is the un-calibrated reading at 0% saturation

References:

Standard Methods for the Examination of Water and Wastewater, 20

th

Ed, 2008. 4500-0 C. Azide

Modification. American Public Health Association. USA.