Page 16: 0 modbus registers, 1 common registers, 2 sensor status registers, 3 device specific measurement registers, Power connections, Communications, Additional information
5.5.2
Power Connections
The red wire provides power for all system modes. Analog output is disabled by default. However,
the 4-20 mA current loop output can be continuous in Modbus or SDI-12 mode as long as Modbus
device register 9507 is set to 1.
5.5.3
Communications
The device automatically switches between Modbus and SDI-12 modes depending on which of the
two interfaces has activity. Modbus and SDI-12 cannot be used at the same time—whichever one is
currently in use will block communication on the other.
5.5.4
Additional Information
For additional information on Modbus and SDI-12 communications, including the SDI-12 commands
and Modbus registers, see the In-Situ Modbus Communication Protocol technical note, available at
6.0 Modbus Registers
6.1
Common Registers
Register
Size
(registers)
Mode & Access Level
(R/W)
Data
Type
Description
9001
1
R/W
ushort Device Id (Device Id = 12 for RDO PRO)
9002
2
R/W
ulong Device Serial Number
9004
3
R/W
time Manufacture Date
6.2
Sensor Status Registers
Register
Size
(registers)
Mode & Access Level
(R/W)
Data
Type
Description
0005
3
R1
time RDO cap start time, 0 = no cap
0008
3
R1
time RDO cap exp. time, 0 = no cap
6.3
Device Specific Measurement Registers
Register
Size
(registers)
Mode & Access Level
(R/W)
Data
Type
Description
Dissolved Oxygen Concentration
0038
2
R1
float Measured value, C
0
0040
1
R1
ushort Parameter Id = 20
0041
1
R1/W2
ushort Units Id
117 = mg/L (default)
118 = µg/L
0042
1
R1
ushort Data Quality Id
0043
2
R1/W3
float Off line sentinel value (default = 0.0)
0045
1
R1
16 bits Available Units = 0x0030 (48)
Temperature
0046
2
R1
float Measured value
0048
1
R1
ushort Parameter Id = 1
Page 16
Modbus Registers