Technical specifications – HT instruments SIRIUS87 User Manual
Page 68
![background image](/manuals/367533/68/background.png)
SIRIUS
87
EN - 66
9. TECHNICAL SPECIFICATIONS
9.1. TECHNICAL
FEATURES
Accuracy is indicated as [% reading + (number of digits) * resolution]. It refers to the
atmospheric conditions listed in § 9.2.1.
- Continuity (LOW
)
Test mode
Measuring range (
)
Resolution (
)
Accuracy
AUTO, R+TIMER, R-TIMER
0,01 – 19,99
0,01
(2% rdg + 2 dgt)
20,0 – 99,9
0,1
Test current
>200mA DC up to 5
(include the resistance of the calibration)
Test current resolution
1mA
Open circuit Test Voltage
6V < V
0
< 12V
- Insulation resistance (R
ISO
)
Test voltage (V)
Measuring range (
)
Resolution (
)
Accuracy
50
0,01 - 19,99
0,01
(2% rdg + 2 dgt)
20,0 - 49,9
0,1
50,0 - 99,9
0,1
(5% rdg + 2 dgt)
100
0,01 - 19,99
0,01
(2% rdg + 2 dgt)
20,0 - 99,9
0,1
100,0 - 199,9
0,1
(5% rdg + 2 dgt)
250
0,01 - 19,99
0,01
(2% rdg + 2 dgt)
20,0 - 199,9
0,1
200 - 249
1
250 - 499
1
(5% rdg + 2 dgt)
500
0,01 - 19,99
0,01
(2% rdg + 2 dgt)
20,0 - 199,9
0,1
200 - 499
1
500 - 999
1
(5% rdg + 2 dgt)
1000
0,01 - 19,99
0,01
(2% rdg + 2 dgt)
20,0 - 199,9
0,1
200 - 999
1
1000 - 1999
1
(5% rdg + 2 dgt)
Automatic selection of the measuring range
Open circuit voltage
Rated voltage test –0% +10%
Short circuit current
<6,0mA with 500V set
Rated measuring current
>2,2mA with 500V on 230k
1mA a 1K
x V
NOM
(
500 V)
- RCD tripping tests (RCD
or RCD
)
RCD type
AC, A General and Selective
Range of phase to earth voltage
100
265V 50Hz + 0.5 Hz
Rated tripping currents (I
N
)
10mA, 30mA, 100mA, 300mA, 500mA
- Contact voltage (U
t
)
Measuring range (V)
Resolution (V)
Accuracy
0
2U
t lim
0,1
-0%, +(5% rdg + 3 dgt)
U
t
lim (U
L
): 25V o 50V