4.3.1.5. Mode

"U

t

"

4.

 Press

the

START/STOP key once: the instrument carries out the

test.


 If the RCD does NOT trip the

instrument emits a

double sound

signal indicating the positive
outcome
of the test and displays the
screen alongside.

RCD

~



1.4

V

OK

N

30

m

A

49

U

t

0° U

L

=

50

V


SAVING:

The tests can be stored pressing the

SAVE key twice

(according to § 5.1).



4.3.2. Tripping times for the general and selective RCDs
If the parameters set on the instrument are in keeping with the type of RCD under test
(and if this works correctly) the test with leakage current I

N

x1, I

N

x2, I

N

x5

SHALL

cause the RCD tripping within the times set according to the following table:

RCD type

I

N

x 1

I

N

x 2

I

N

x 5

Description

General 0.3s

0.15s

0.04s

Max tripping time in seconds

Selective

S

0.5s 0.20s 0.15s

Max tripping time in seconds

0.13s 0.05s 0.05s

Minimum delay tripping time in
seconds

* For rated values I

N  30mA the test current at five times is 0.25A.

For currents equal to ½

I

N

the RCD shall not trip in any case.

Table 3: Tripping times for tests with leakage currents I

N

x1, I

N

x2, I

N

x5 and AUTO

Value of earth resistance

R

a

. The display of the

message "

o.r." means that the instrument detected

a resistance higher than 1999

.

Value of contact voltage

U

t

detected referred to the

rated value of the RCD current set.

It indicates if the
test was carried
out with current
phase 0° or 180°
with respect to
the voltage.

SAVE

OK: indicates that
the RCD passed
the test.

or RCD

START

STOP