I-v400 - solar i-v, En - 57, List of display messages – HT instruments SOLAR I-V User Manual
Page 59

I-V400 - SOLAR I-V
EN - 57
6.4. LIST OF DISPLAY MESSAGES
MESSAGE DESCRIPTION
Voltage too low
Check the voltage between C1 and C2 input terminals
Vin > 1000
DC output voltage from module/string > 1000V
Irradiance too low
Irradiance value lower than threshold limit
NTC Error
Internal NTC damaged. Contact service dpt
Please wait for cooling…
Instrument overheated. Wait before starting the tests again
Memory full
Internal memory full. Download data to PC
Pulse width too long
Anomalous condition. Repeat test with more modules
Current too low
Measured current lower than the minimum detectable
Vdc wrong connection
Check the voltage between C1 and C2 input terminals
Negative voltage
Check the polarity of input terminals
Data base full
The number of modules defined is > 30
Data @ STC unavailable
The meter does not calculate the date at STC conditions
Irradiance too high
Irradiance value higher than the maximum range
Data unavailable
Generic error. Repeat the test
Isc too high
Output current higher than maximum range
Wrong date
Set a correct date/hour on meter
Error 1/2/3/4/5: contact Assistance
Contact service dpt
Error EEPROM : contact Assistance
Contact service dpt
Error FLASH : contact Assistance
Contact service dpt
Error RTC : contact Assistance
Contact service dpt
Battery low
Low battery indication. Fit new batteries inside meter
Error: Vmpp >= Voc
Check the settings of module inside DB
Error: Impp >= Isc
Check the settings of module inside DB
Error: Vmpp * Impp >= Pmax
Check the settings of module inside DB
Error: alpha too high
Check the settings of module inside DB
Error: beta too high
Check the settings of module inside DB
Error: Toll too high
Check the settings of module inside DB
Error: gamma too high
Check the settings of module inside DB
Module already present
Name of module just used inside DB
Delta-Irrad. too high. Retry
Unstable condition on irradiance. Repeat the test
Voltage not steady
Anomalous condition. Repeat test with more modules
Unsteady Current
Difference between 2 consec. Instant. values of current is > 0.13A
Firmware mismatch
Problem with internal FW. Contact service dpt
Ref. Cell temp over range
Temperature measured by reference cell is too high
PV module temp over range
Temperature of module over maximum range
Wrong Mod. Num. Continue?
Setting of number of modules not ok with measured Voc
Ref. Cell temp not detected (ENTER/ESC)
Measure on the cell of module not performed
Unable to calculate Rs value
Rs value external to measurement range
Thermal Instability
The instrument detect unsteady conditions (see § 10.2)
Error during memory writing
Problem with the access to internal memory
Error Radio: contact Assistance
Contact service dpt
Error Radio transmission
Contact service dpt
Error during download
Contact service dpt
Recording downloaded
Recording corrected downloaded on the instrument
Radio connection activated
Corrected RF connection with remote unit SOLAR-02
Wait data analysis
Download data from SOLAR-02 and wait for response
Unable execute analysis
Problem on downloaded data from SOLAR-02. Verify settings
I < Lim
Output current lower that the minimum measurable
CAUTION: internal short
Contact service dpt
Remote unit not detected Enter/Esc
Unit SOLAR-02 do not RF connected to the instrument
Rem. Unit in rec
Unit SOLAR-02 in recording parameters Irr/Temp
STC Data available after Stop rec.
Stop the recording to obtain STC data
STC Data available in MEM
Data STC condition save on memory
Verify connect. P1
Check the proper insertion of test cable in the input P1
Error in data download from SOLAR-02
SOLAR I-V set for use with MPP300. Error in download from SOLAR-02.
SOLAR-02 not detected. Stop Rec.?
SOLAR I-V set for use with MPP300. SOLAR-02 not detected when stopping Rec
MPP300 not detected
SOLAR I-V set for use together with MPP300. MPP300 not detected
SOLAR-02 not detected.
SOLAR I-V set for use together with MPP300. SOLAR-02 not detected.
MPP300: negative AC power
MPP300 has detected negative AC powers.
MPP300: inverted AC and DC voltages
MPP300 has detected inverted AC and DC voltages
No power supply to MPP. Continue? Enter/Esc
MPP300 has not detected the presence of the external power supply.