GW Instek GSP-827 Programming Manual User Manual
Page 10
Copyright 2003 GOODWILL INSTRUMENT CO., LTD.
CONtrol RBW
MANual
200
Assign the RBW in manual mode
CON:RBW:MAN
OK
CONtrol RBW
MODE?
200
Query the RBW mode.
Auto, Manual
CON:RBW:MODE?
Manual
CONtrol RBW?
200
Query the RBW
CON:RBW?
30 kHz
CONtrol VBW
400
Assign the VBW by index number
0=10Hz, 1=30Hz, 2=100Hz, 3=300Hz, 4=1kHz,
5=3kHz, 6=10kHz, 7=30kHz, 8=100kHz,
9=300kHz, 10=1MHz
CON:VBW^4
OK
CONtrol VBW
AUTO
400
Assign the VBW in auto mode
CON:VBW:AUTO
OK
CONtrol VBW
MANual
200
Assign the VBW in manual mode
CON:VBW:MAN
OK
CONtrol VBW
MODE?
700
Query the VBW mode.
CON:VBW:MODE?
Manual
CONtrol VBW?
600
Query the VBW
CON:VBW?
1 Khz
CONtrol SYStem GPIB
ADDRess
300
Assign the GPIB address
1 to 30
CON:SYS:GPIB:ADDR^9
OK
CONtrol SYStem GPIB
ADDRess?
500
Query the GPIB address
CON:SYS:GPIB:ADDR?
9
CONtrol SYStem CONTrast
200
Assign the contrast number of the
LCD module
128(white)~176(black)
CON:SYS:CONT^145
OK
CONtrol SYStem CONTrast?
400
Query the contrast number of LCD
module.
CON:SYS:CONT?
145
CONtrol SYStem CLOCK?
100
Query the system date and clock
CONtrol:SYStem:CLOCK?
23:53:00
Monday May 26, 2003
CONtrol SYStem FirmWare?
700
Query the Firmware version
CONtrol:SYStem:FirmWare?
Firmware version = 2.0
CONtrol SYStem VERsion?
600
Query the Software version and built
date
CONtrol:SYStem:VERsion?
System Software Version = 00.94
03/31/03
CONtrol SYStem PRESET
200
Preset the system
CON:SYS:PRESET
OK
CONtrol SYStem DISPlay
100
Set the display mode. 0:Normal,
1:Inverse
CON:SYS:DISP^0
OK