INTRODUCTION

7

5. Warning

Indicator

The warning indicator lights up
when high voltages are produced
from terminals T1, T2 or T3. The
warning indicator will flash when
in standby mode.

6. Measuring

Terminals

Measuring Terminals T1 and T2 are
used to measure leakage current.
Terminal T2 has a replaceable fuse
(250V, 32mA).

7. Circuit

Breaker

The circuit breaker has over-current
protection for the EUT rated at 15A.
When testing, the warning indicator
will illuminate.
I: ON, normal operation
O: OFF, inactive or during over-
current protection.

8. EUT AC

Terminal
Block

Supplies AC power for the EUT.
Includes automatic shut-down
(circuit breaker) with over-current
protection. Maximum current
output 10A, maximum power
output, 1500VA.

9. T3 110%

Voltage
Application

An isolated voltage (1:1) is output to
T3 from the EUT AC IN voltage by
an isolation transformer. This
terminal is limited to medical
networks (MD:F)

10. USB HOST

USB host terminal.