GCP Series User Manual
36
Common specifications (for GCP-530 and GCP-1030)
GCP-530
GCP-1030
Operating
temperature and
humidity range
0 to 40°C (32 to 104°F), 80 % RH or less (no
condensation)
Rated Supply
voltage
12 V 0.5 V
Maximum Peak
Current Input
Non-continuous 50 Apeak
Maximum
Continuous Input
Range
30 Arms (Derating according to frequency shown in Fig. 3A
and 3B)
Output Voltage
Rate
0.1 V/A
Amplitude Accuracy +1.0% rdg. ±1 mV ; up to 30 Arms
+2.0% rdg. ; over 30 Arms to 50 Apeak (DC, and 45 to 66
Hz, input within continuous maximum input range)
Noise
Equivalent to 2.5 mArms or less (for 20 MHz band
measuring instrument)
Input Impedance
(Characteristics shown in Fig.4A and 4B)
Temperature
Coefficient for
Sensitivity
±2% or less (During input of 50 Hz 30 Arms within range of
0 to 40°C (32 to104°F)
Storage
Temperature and
Humidity Range
-10 to 50°C (14 to 122°F), 80 % RH or less (no
condensation)
Location for Use
Indoor, altitude up to 2000 m (6562 feet)
Effect of External
Magnetic Fields
Equivalent to a maximum of 20 mA (DC and 60 Hz,
Magnetic field of 400 A/m)
Maximum Rated
Voltage
300 V, CAT Ⅰ(insulated conductor)
Diameter of
Measurable
Conductors
5 mm dia. 0.2" dia.
Guaranteed
Accuracy Period
1 year (Opening/closing up to 10,000 times)
Cable Lengths
Sensor cable Approx. 1.5 m (59.0"),
Power supply cable Approx. 1 m (39.4")