GCP Series User Manual

36

Common specifications (for GCP-530 and GCP-1030)

GCP-530

GCP-1030


Operating
temperature and
humidity range

0 to 40°C (32 to 104°F), 80 % RH or less (no
condensation)

Rated Supply
voltage

12 V 0.5 V

Maximum Peak
Current Input

Non-continuous 50 Apeak

Maximum
Continuous Input
Range

30 Arms (Derating according to frequency shown in Fig. 3A
and 3B)

Output Voltage
Rate

0.1 V/A

Amplitude Accuracy +1.0% rdg. ±1 mV ; up to 30 Arms

+2.0% rdg. ; over 30 Arms to 50 Apeak (DC, and 45 to 66
Hz, input within continuous maximum input range)

Noise

Equivalent to 2.5 mArms or less (for 20 MHz band
measuring instrument)

Input Impedance

(Characteristics shown in Fig.4A and 4B)

Temperature
Coefficient for
Sensitivity

±2% or less (During input of 50 Hz 30 Arms within range of
0 to 40°C (32 to104°F)

Storage
Temperature and
Humidity Range

-10 to 50°C (14 to 122°F), 80 % RH or less (no
condensation)

Location for Use

Indoor, altitude up to 2000 m (6562 feet)

Effect of External
Magnetic Fields

Equivalent to a maximum of 20 mA (DC and 60 Hz,
Magnetic field of 400 A/m)

Maximum Rated
Voltage

300 V, CAT Ⅰ(insulated conductor)

Diameter of
Measurable
Conductors

5 mm dia. 0.2" dia.

Guaranteed
Accuracy Period

1 year (Opening/closing up to 10,000 times)

Cable Lengths

Sensor cable Approx. 1.5 m (59.0"),
Power supply cable Approx. 1 m (39.4")