9 view channel history, View channel history, Figure 26 set-up 4-20ma outputs – Detcon 6400-N4X User Manual
Page 23: Figure 27 view channel history

1600/6400-N4X
1600/6400-N4X- Instruction Manual
Rev. 1.1
Page 19 of 38
Figure 26 Set-up 4-20mA Outputs
For example, assume setting up two AO-4 modules. Make sure to set each AO-4 module to a unique ID#
using the front panel switches. If Device #’s 1 through 4 are already being used for DA-4 and RL-4 modules,
then set these two AO-4 modules to 5 and 6. The entered values should then look like this:
CH#
Device
Address
01
05
01
02
05
02
03
05
03
04
05
04
05
06
01
06
06
02
07
06
03
08
06
04
5.3.9 View
Channel
History
View Channel History controls all of the logged history of gas readings for each active channel. Select the
channel number of interest and use the scroll arrows to review that channels data. The history log includes the
Date/Time, Time-Weighted-Average (TWA), and the Peak Reading for each hour of operation. When the
memory buffer is full, the oldest data is replaced. The log can retain data for at least 30 days before the buffer
is full and the oldest data is replaced.
Figure 27 View Channel History