Range of Detection
The menu item appears as: “Range XXX”
Serial ID address.
The menu item appears as: “Serial ID XX”
AutoSpan Level.
The menu item appears as: “Auto Span Level XX”
Days From Last AutoSpan
The menu items appears as: “Last Cal XX days”
Remaining Sensor Life
The menu item appears as: “Sensor Life 100%”
Sensor Bridge Current
The menu item appears as: “Bridge XXXmA
Sensor Bridge Voltage
The menu item appears as: “Bridge X.XXVDC
Gas Factor
The menu item appears as: “Gas Factor X.X”
Cal Factor
The menu item appears as: “Cal Factor X.X”
4-20mA Output
The menu item appears as: “mA Output X.XXmA”
Input Voltage Supply
The menu item appears as: “Voltage XX.XVDC”
Operating Temperature
The menu item appears as: “Temp= XX C”
When the status list sequence is complete, the ITM will revert to the “View Sensor Status” text scroll. The
user can either: 1) review list again by executing another 3-4 second hold, 2) move to another menu item by
executing a momentary hold over PGM1 or PGM2, or 3) return to Normal Operation via automatic timeout of
about 15 seconds (the display will scroll “View Sensor Status” 4 times and then return to Normal Operation).
3.5.3
Set AutoSpan Level
Set AutoSpan Level is used to set the span gas concentration level that is being used to calibrate the sensor.
This level is adjustable from 5% to 95% (0-100% LEL range) or 2% to 50% (0-50% LEL range). The current
setting can be viewed in View Program Status.