DeFelsko PosiTector 200 v.3.1 Full Manual User Manual
Page 14
14
Graphics
ü
(Advanced models only)
Gage Info
Displays the model number & serial number, probe type & serial
number, PosiSoft.net registration key, the amount of remaining
memory for storage of readings, date and time and software
packages.
For security purposes, the registration key is required to add the
Gage to your PosiSoft.net account.
When selected, the Gage displays a
graphical representation of the
ultrasonic pulse as it travels through the
coating system.
As the probe is depressed and the
ultrasonic pulse travels through the
coating system, the pulse encounters
changes in density at the interfaces
between coating layers and between the
coating and the substrate.
These interfaces are depicted by a "peak". The greater the change
in density, the higher the peak. The more gradual the change in
density, the greater the width of the peak. For example, two coating
layers made of essentially the same material and "blended" would
result in a low, wide peak. Two materials of very different density
and a well-defined interface would result in a high, narrow peak.
The PosiTector 200 chooses the highest peak (single layer) or
peaks (multi-layer) within the Set Range. For example, if the
number of layers is set to 3, the three highest peaks would be
identified with small red triangles. The peak values are also
displayed numerically as thickness measurements.
The Graphics display can also be accessed using the
Set Range menu option.