Setup menu – DeFelsko PosiTector 200 v.3.1 Full Manual User Manual
Page 13
13
Reset
Reset (soft reset) restores factory settings and returns the Gage
to a known condition. The following occurs:
- All batches, stored measurements, images, batch names and
screen captures are erased.
- All calibration and range adjustments are cleared and returned
to the Gage’s factory settings.
- The icon will appear on the display.
- Menu settings are returned to the following:
Perform a more thorough Hard Reset by powering down the
Gage, waiting several seconds, then simultaneously holding both
the center and (+) buttons until the Reset symbol
appears. This returns the Gage to a known, “out-of-the-box”
condition. It performs the same function as a menu Reset with the
addition of:
- Bluetooth Pairing info is cleared.
- Menu settings are returned to the following:
Setup Menu
Setup Menu
Units = microns
Flip Display = Normal
White on Black = OFF
Language = English
Battery Type = Alkaline
Backlight = Normal
USB Drive= ON
Auto Sync = OFF
Polyurea = OFF
Memory = OFF
Graphics = ON
Cal Lock = OFF
Bluetooth = OFF
Display = None
Layers = 1
Layer Name = Layer 1
Layer Color = blue
Date, Time and WiFi settings are not affected by either
Reset.
NOTE: