beautypg.com

COSA Xentaur COSA 707 User Manual

Page 40

background image

COSA Instruments

USER MANUAL COSA 707

OK

OK

,

change barometric pressure / area

F2

change units

F1

measurement START / STOP

F3

mode auto: automatic input of T gas (adoption measured T gas value)

PRINTER

print-out actuel windows display

OK or ESC back to measurement

The Pitot tube shaft is introduced perpendicularly in to the duct, by holes made in the surface, at
selected positions.(see illustration).

The probe with the ellipsoidal nose is held parallel to and against the flow. The total pressure (+)
collected by the nose is connected to the (+) connection on the COSA 707.

The static pressure (-) collected by the static pressure intakes in the probe is connected to the -
connection on the COSA 707.

The COSA 707 will then indicate the dynamic pressure often called the velocity pressure.

The dynamic pressure corresponds to the difference between the total pressure and the static
pressure.

P dyn. = P tot. + P stat.

The calculation of the gas flow velocity occurs by following formula:

Where :

P stat << P baro