COSA Xentaur COSA 707 User Manual
Page 22
COSA Instruments
USER MANUAL COSA 707
www.cosaxentaur.com
22
9.4 Configuration of measurement program
(Flue gas measurements) Select one of the 4 configurable measurement programs.
For each of the programs the following parameters can be configured:
•
CO ppm limit: adjustable value for the CO sensor protection. If the CO value in the flue gas
is higher than the adjusted value in the analyzer, the purge pump will be activated and the
sensor will be protected against high CO concentrations. (Optionalalal)
•
Selectable fuel types: choose and select from the available fuel type list
•
Measurement windows: configuration of what and where will be displayed in the 3 meas-
urement value windows.
•
Zoom – window: select what will be displayed in the zoom - window
•
Program name
The following programs have a pre configured:
•
“Test Program” indented for use at instrument maintenance and calibration
9.4.1 Setting the CO ppm limit values
The CO ppm limit can be adjusted in the window „Measuring program selection“.
Select one of the available programs (arrow up/down) then press the F1 key.
F1
,
,
Select one available program
F1
Open the CO ppm limit window
The CO-ppm limit value can be adjusted in 100 ppm
steps between 300 ppm and 4.000 ppm / 10.000
ppm
)
OK oder ESC
Return to the measurement window