Draft and temperature are measured at the tip of the sampling probe.

5.1 Gas flow diagram

Gas flow diagram with CO purging pump

Position Description

1

Sampling probe

2

Triple hose

3

Condensate separator

4

Star filter

5

Non return valve

6

Gas pump

7

Sensor chamber

8

O

2

-Sensor

9

2. sensor

10

3. sensor

11

4. sensor

12

5. sensor

13

CO Purge pump

14

Pressure sensor *

* optional

5.2 Electrochemical measured principle

The oxygen content of the sample gas is measured with a 2 electrode electrochemical
sensor.

Toxic gases like carbon monoxide (CO), nitrogen oxide (NO), nitrogen dioxide (NO

2

), sul-

phur dioxide (SO

2

), and hydrogen sulphide (H

2

S) are measured with 3 electrode sensors.

The electrochemical sensors are based on gas diffusion technology.