Draft and temperature are measured at the tip of the sampling probe.
5.1 Gas flow diagram
Gas flow diagram with CO purging pump
Position Description
1
Sampling probe
2
Triple hose
3
Condensate separator
4
Star filter
5
Non return valve
6
Gas pump
7
Sensor chamber
8
O
2
-Sensor
9
2. sensor
10
3. sensor
11
4. sensor
12
5. sensor
13
CO Purge pump
14
Pressure sensor *
* optional
5.2 Electrochemical measured principle
The oxygen content of the sample gas is measured with a 2 electrode electrochemical
sensor.
Toxic gases like carbon monoxide (CO), nitrogen oxide (NO), nitrogen dioxide (NO
2
), sul-
phur dioxide (SO
2
), and hydrogen sulphide (H
2
S) are measured with 3 electrode sensors.
The electrochemical sensors are based on gas diffusion technology.