beautypg.com

2 introduction, 1 appropriate use, 1 application – LumaSense Technologies IS 12-Al/S User Manual

Page 9: 2 scope of delivery, 3 technical data

background image

IS 12 AI & IS 12-AI/S

Introduction 9

2 Introduction

2.1 Appropriate use

The IS 12-Al is a special version of the pyrometer IS 12, designed for the measurement of
aluminum. The robust die-cast housing with protection class IP65 is designed for the use in

industrial environments.

The instruments IS 12-Al/S is equipped with an integrated scanner which moves the measuring
beam adjustable up and down up to 4°.

2.1.1 Application

The IS 12-Al is a special development of the well proven IS 10-Al for
aluminum applications in temperature ranges between 350 and
1050 °C.

Conventional pyrometers – even 2-color pyrometers – are not able to

measure the temperature of aluminum correctly due to the special
physical properties of aluminum. To avoid the negative influence of
these properties in the non-contact temperature measurement the
IS 12-Al operates in a special spectral range. In this spectral range solid
aluminum has a very high and stable emissivity between 30 and 43%.
That is why the instrument is suitable for aluminum applications such

as extrusion, rolling, billet heating and other heating processes.

Note: Due to physical reasons the IS 12-Al is sensitive to light at the beginning of the
temperature range. This sensitivity decreases with increasing temperature. In
applications with low temperatures (at the start of the temperature range) the
measuring area has to be screened against light to avoid its influence (see drawing).

2.2 Scope of Delivery

Instrument with selectable optics, PC measurement and evaluation software InfraWin, works

certificate, and operation manual.

2.3 Technical Data

Temperature ranges:

350 to 900 °C

(MB 9)

400 to 1050 °C (MB 10.5)

Sub range:

any range adjustable within the temperature range, minimum span 51 °C

Signal processing:

photoelectric current, digitized

Spectral ranges:

narrow band in the near infrared

Power supply:

24 V DC (15 to 40 V DC) or 24 V AC (12 to 30 V AC), 48 to 62 Hz

Power consumption:

max. 7 W

Analog output:

0 to 20 mA or 4 to 20 mA switchable, load 0 to 500 Ohm

Test current output:

10 mA, fixed

Digital interface:

switchable: RS232 or RS485 addressable (half duplex),
baud rate 2.4 up to 115.2 kBd

Display:

built-in 5 digit LED display, height 13 mm, additional function LED’s

This manual is related to the following products: