beautypg.com

LumaSense Technologies IS 12-Al/S User Manual

Page 34

background image

IS 12 AI & IS 12-AI/S

Data format UPP (Universal Pyrometer Protocol) 34

Description

Command

Parameters

Limit contact 1:

AAs1XXXX

XXXX = switch point limit contact 1

(ASCII-hex, 4 digit, full degree in °C or °F)

Limit contact 2:

AAs2XXXX

XXXX = switch point limit contact 2

(ASCII-hex, 4 digit, full degree in °C or °F)

Setting hysteresis of limit

contacts:

AAhlXX

XX = 2 … 20, full degree in °C or °F

Changing °C / °F

AAfhX

Output: X = 0: display in °C; X = 1: display in °F

interface type:

AAin

Output: X = 1 or 2; 1 = RS232, 2 = RS485

Changing address:

AAgaXX

XX = (00 ... 97)

00 ... 97 = regular device addresses

99 = Global address with response

98 = Global address without response (only setting commands!)

Changing

baud

rate:

AAbrX

X=0...6 or 8 (dez.)

0 = 1200 baud

3 = 9600 baud

6 = 57600 baud

1 = 2400 baud

4 = 19200 baud

(7 is not allowed)

2 = 4800 baud

5 = 38400 baud

8 = 115200 baud

Wait time:

AAtwXX

XX = 00 ... 99 (decimal)

Error status:

AAfs

Output 1 byte hex (00 = no error)

Bit 0 = 1: Measurement unit doesn’t work

Bit 1 = 1: Internal temperature measurement doesn’t work

Lock keyboard:

AAlkX

X = 0 ... 3

1 = lock lk1, removal with command lk0 or power off-on

0 = removal of lock lk1

3 = continuous lock lk3, removal only with command lk2

2 = removal of lock lk3

Reading

parameters:

AApa

Output decimal 11-digit:

Digit 1 und 2 (10...99 or 00): e

Digit 3

(0

...

6):

t

90

(Exposition time)

Digit 4

(0

...

8):

t

CL

(max. storage clear mode)

Digit 5 (0 ... 1): Analog output

Digit 6 und 7: (00 ... 98): Temperature

Digit 8 und 9 (00 ... 97): Address

Digit 10 (0 ... 6 or 8): Baud rate

Digit

11

(0):

always

0

Laser

targeting

light:

AAlaX

X = 0 switch off laser

X = 1 switch on laser

Internal

temperature:

AAgt

Output: XX (dez. 00 ... 98, in °C)

XXX

(dez.

032

...

208

°F)

Max. internal

temperature:

AAtm

Output: XX (dez. 00 ... 98, in °C)

XXX

(dez.

032

...

208

°F)

Ref. number:

AAbn

Output: XXXXXX (hex 6-digit)

Serial number:

AAsn

Output: XXXX (hex 4-digit)

Device type:

AAna

Output: “IS 12-Al” or “IS 12-Al/S” (16 ASCII-characters)

Device type /

software version:

AAve

Output: XXYYZZ (6-digit decimal)

XX = 07 (IS 12-Al and S 12-AL/S)

YY = Month of software version

ZZ = Year of software version

Software version in detail: AAvs

tt.mm.yy XX.YY

tt = Day; mm = month; yy = year; XX.YY = software version

Note: the letter ”l” means the small type of “L”