All instrument settings can be done directly at the instrument. To activate the adjusting keys, a
tip of ball pen must be used. This avoids changing parameters by mistake.
4.1 Key panel operation
4.2 Functions and setting of parameters
Temperature or
parameter display
In the measuring mode the display shows the actual temperature
reading. After pushing the parameter key the display indicates the actual
value of one selected parameter.
Special indications: 8888 = measurement exceeds the end of sub range
Targeting light
on / off
Pushing the PL key in measuring mode the laser
targeting light is switched on. The IS 12-AI measures
continuously also with switched on targeting light and
the display shows the actual temperature reading. The
laser targeting light is switched off automatically after
approx. 2 min or after a further push of the PL key.
Setting keys 1) PAR
2)
With the PAR button all available parameters are displayed in the
following description. Pushing the button again changes the display to
the next parameter, after the last parameter it changes to the actual
temperature reading.
With the arrow keys and
all parameter settings can be displayed.
Pushing the button longer changes the settings in fast mode (the keys
and ENT are blocked if the keyboard is locked via the interface
commands or/and during a measurement in combination with a PC and
InfraWin).