Data format UPP (Universal Pyrometer Protocol) 33
9 Data format UPP (Universal Pyrometer
Protocol)
Via interface and a suitable communication software or via “Test” function of the InfraWin
commands can be exchanged directly with the pyrometer.
The data exchange occurs in ASCII format with the following transmission parameters:
The data format is: 8 data bits, 1 stop bit, even parity (8,1,e).
The device responds to the entry of a command with: output (e.g. the measuring value) + CR
(Carriage Return, ASCII 13), to pure entry commands with “ok” + CR.
Every command starts with the 2-digit device address AA (e.g. “00”). This is followed by 2 small
command letters (e.g. “em” for level of emissivity
e
), finished with CR
This is followed, if necessary for that command, by the ASCII parameter “X”. If this parameter
“X” is omitted, then the device resets with the current parameter.
A „?“ after the small command letters answers with the respective settings (only at setting
commands, not at enquiry commands).
Example: Entry: “00em“ + CR
The emissivity setting (
e
) of the device with the address 00 is returned
Answer: “0970“ + CR means Emissivity = 0.97 or 97.0%
Description
Command
Parameters
Reading temperature
value:
AAms
Output: XXXXX (dec., in
1
/
10
°C or °F)
last digit is the decimal place (88880 = Temp.-Overflow)
Reading temperature
value repeated:
AAmsXXX
XXX = 000…999 (XXX = number of measuring values)
Reading basic
temperature range:
AAmb
Output: XXXXYYYY (hex 8-digit, °C or °F)
XXXX
=
beginning
of
temp.
range
YYYY=
end
of
temp.
range
Reading temperature sub
range:
AAme
same as mb
Reading emissivity:
AAem
Output: XXXX (dez. 0010 … 1000 in ‰)
Emissivity:
AAemXX
XX = (10…99%), 00 = 100% (decimal)
Emissivity:
AAemXXXX
XXXX = (0010 … 1000‰) (decimal)
Exposition time t
90
:
AAezX
X = 0 … 6 (decimal)
0 = intrinsic time constant
1 = 0.01 s
3 = 0.25 s
5 = 3.00 s
2 = 0.05 s
4 = 1.00 s
6 = 10.00 s
External deletion:
AAlx
Simulation of an external deletion contact
Delete
maximum value: AAlzX
X = 0 … 8 (dez.)
0 = Maximum value storage off
1
=
0.01
s
4
=
1.00
s
7
=
external
deletion
2 = 0.05
s
5 =
5.00 s
8 = automatically deletion
3 = 0.25 s
6 = 25.00 s
Setting of temperature
sub range:
AAm1XXXXYYYY
XXXX (hex 4-digit) beginning of temp. range (°C)
YYYY (hex 4-digit) end of temp. range (°C)
Analog output:
AAasX
X = 0 or 1; 0 = 0 to 20 mA
1 = 4 to 20 mA