IGA 6/23 Advanced Manual

Settings / parameter descriptions · 25

approximately 0.5 ms. The response time can be extended to 1 ms; 3 ms; 5 ms; 10 ms; 50 ms;
250 ms;1 s; 3 s; 10 s.

Note: Settings for Response Time t

90

min, 1 ms; 3 ms; 5 ms; 10 ms; 50 ms; 250 ms; 1 s; 3 s; 10 s

4.6 Clear Peak Memory (t

CLEAR

)

The integrated maximum value storage is activated when the parameter t

clear

is set to something other than “OFF” or “HOLD”.
If the maximum value storage is switched on, the highest last temperature
value will always be displayed and stored. As such, it may be beneficial to
periodically clear and reset the stored maximum values in order to obtain

new temperature readings.
This storage also has to be cleared at regular intervals when fluctuating
object temperatures cause the display or the analog outputs to change too
rapidly or when the pyrometer is not constantly viewing an object to be
measured.

4.6.1 Single and Double Storage Modes

Depending upon the selected settings, the maximum value storage will either work in single

storage mode or in double storage mode.
Single Storage

Mode:

Single storage mode is used when you want to reset the stored value using
an external impulse via one contact closure from an external relay (such as
between two measured objects). The relay contact is connected directly to
the pyrometer between pins J and K. This mode allows a new value to be
established after each impulse from the reset signal. Single storage mode

also comes into effect when the Clear Peak Memory t

clear

is set to AUTO.

Double Storage

Mode:

Double storage mode comes into effect when selecting one of the reset
intervals. This mode utilizes two memories. With the first memory, the

highest measured value is held and is deleted alternately in the time
interval set (clear time). The other memory retains the maximum value
throughout the next time interval. The disadvantages of fluctuations in the
display with the clock frequency are thereby eliminated.

Note: The maximum value storage setting coincides with adjustments made to the

response time.
The response time setting (working like a low-pass filter) is applied first. After that,
the maximum storage is processed. So when using both, the maximum storage takes
the peak of the signal that was previously smoothed by the response time filter.

Settings:

off

10 ms
50 ms

250ms

1 s
5 s

25 s

extern

auto