The flange system is a modular mounting system to fix the pyrometer on furnaces, vacuum
chambers, etc.

It can consist of e.g. mounting support, tube support with air purge and flange and an open or

closed ceramic sighting tube. The mounting support can be equipped with a quartz window for
vacuum applicationsIt may consist of an equipment rack, flange, and an open or closed ceramic
tube. The equipment rack can be equipped for vacuum applications with a fused silica.

2.7 Transport, Packing, Storage

With faulty shipping, the instrument can be damaged or destroyed. To transport or store the
instrument, please use the original box or a box padded with sufficient shock-absorbing
material. For storage in humid areas or shipment overseas, the device should be placed in
welded foil (ideally along with silica gel) to protect it from humidity.
The pyrometer is designed for a storage temperature of -20 to 80 °C with non-condensing
conditions. Storing the instrument out of these conditions can cause damage or result in
malfunction of the pyrometer.

Schematic drawing of the flange system

Pyrometer

IGA 6/23

Advanced

Ceremaic

sighting tube

600 x 24, closed

Instrument’s

support