beautypg.com

Yokogawa µR20000 User Manual

Page 39

background image

4-8

SM 04P01B01-01E

Procedure

1.

Press the

MENU

and

CH UP

keys while turning ON the power. Menu = A/D Adjust

is displayed.

2.

With Menu = A/D Adjust is displayed, press the

key.

Menu=A/D Adjust

3. Input the password

(see page 3-6).

Password=

4.

Select Mode = Cal/Exec, then press the

key.

Mode=Cal/Exec

5.

Select A/D No. = 1, then press the

key.

A/D No.=1

6.

Select Range = 20 mV.

Range=20mV

7.

Enter the zero value for the 20 mV range on channel 1 and the Full value on
channel 2, then press the

key.

After measurement, the adjusted value is displayed.

Range=20mV
Z:-585 F:32769

8.

Press the

key.

You are returned to the A/D No. = screen.

9.

Repeat steps 5 to 8 to adjust the following ranges: 60 mV, 200 mV, 1 V, 2 V, 6
V, 20 V, Pt100, Cu10 (with the /N1 option).

Saving Adjusted Values
10.

When all adjustments are finished, select Mode = End, then press the

key.

Mode=End

11.

Select A/D No. = 1, then press the

key.

A/D No.=1

12.

With End = Store displayed, press the

key.

The adjusted value is saved to the A/D converter.

End=Store

Concluding Adjustments
13.

Select Menu = End, then press the

key. You are returned to the operation

mode.

Menu=End

4.3 Adjustments to the Dot Model

* Select using the key. To enter numerical values, move the cursor to the entry point using the
key. Select the value using the key. You can hold down the SHIFT key while pressing
key or key to reverse the key’s action.