2 self diagnostic test, Self diagnostic test -2 – Yokogawa µR20000 User Manual
Page 15
3-2
SM 04P01B01-01E
3.2
Self Diagnostic Test
The recorder is provided with complete self diagnostic functions to enhance reliability in
measurement and serviceability.
When you turn ON the power, the recorder will automatically execute the following types
of diagnoses alternately and display the results. After these tests are completed, the
recorder is ready for use.
1.
Main program (Flash ROM) test
2.
Main RAM write/read test
3.
A/D and A/D EEPROM test
4.
Memory acquisition test (write test from Flash)
5.
Mechanical system test
6.
Ethernet module test
The table below shows the results of the self diagnostic tests when a problem is
detected.
Error display
Possible Cause
B3
secondary copy error.
Problem with CPU Assembly
B4
Flash Device error.
B5
Flash SUM error.
B6
Flash Data error.
902 RAM
failure.
910
A/D error.
Problem around the A/D Board PCB or EEPROM
951
EEPROM write error.
921
A/D calibration value error.
Problem with the A/D Board PCB or wrong adjustment data value for
922
A/D calibration is in the wrong order.
the A/D Board PCB.
930
Memory acquisition failure.
Problem with CPU Assembly
940
The ethernet module is down.
Problem with the COMM PCB or CPU Assembly
950
A/D number error.
Problem with the A/D Board PCB (installation of the A/D Board PCB
out of spec)
960 Ribbon
error.
(Dot model) Problem with the ribbon shift mechanism or Dot10 CPU
Assembly
961 Printer
error.
(Dot model) Abnormal movement of the printing carriage or problem
with the Dot10 CPU Assembly
962 Plotter
error.
(Pen model) Problem with Pen CPU Assembly
963
Pen 1 error.
(Pen model) Abnormal movement of the Pen Servo Assembly or
964
Pen 2 error.
problem with the Pen CPU Assembly
965
Pen 3 error.
966
Pen 4 error.