3 performance tests, Test environment, Test instruments – Yokogawa µR20000 User Manual
Page 16: Performance tests -3, Test environment -3 test instruments -3, Warning, Caution
3-3
SM 04P01B01-01E
Testing
3
3.3
Performance Tests
Read the warning and cautions below before beginning tests.
WARNING
• Circuit patterns of the printed circuit board are exposed. Be careful when
handling so that hands or fingers are not injured by the protruding pins.
CAUTION
• Circuit patterns of the printed circuit board are exposed. If other metallic
materials touch these patterns, electrical shorting will occur, causing the circuit
to be damaged or burnt.
• It is sometimes necessary to turn the recorder upside down for adjustment. Do
not drop the instrument, or allow it to fall over.
Test Environment
Item
Conditions
Ambient temperature
23 ± 5°C
Relative humidity
55 ± 20%
Atmospheric pressure
86 kPa-106 kPa
Power supply voltage
Nominal voltage ±1%
Power supply frequency
Nominal frequency ±1%
Power supply waveform
Distortion factor of 5% or less
DC power supply ripple
Content ratio of 0.1% or less
Vibration
A value having negligible effects on the instrument
Position
Within ±3% of the specified position
Interference*
A value that does not affect the measured results
Electrical field
A value that does not affect the measured results
Magnetic field
A value that does not affect the measured results
Atmospheric pollutants
Levels of corrosive gasses, vapors, salts, and dust that
do not affect the measured results.
Other external effects**
A value that does not affect the measured results
*
Interference
refers to common mode noise, series mode noise, power supply noise, and other
phenomena in the signal line.
** Other external effects include luminance, draft, ultrasonic waves, and radiation.
Test Instruments
Instrument
Specifications
DC voltage generator
Accuracy: ±0.01% of setting or better
DMM
Accuracy: ±0.01% of rdg or better
Variable resistors
Accuracy: 0.01% + 2 mΩ or better
Insulation tester
500 VDC
Withstand voltage tester
1 to 1.5 kVAC
Glass scale
105 mm or more