Key Operations on the Recorder

The recorder has the following modes. The following is an explanation of how to enter
and exit each mode.
• Operation Mode

The mode in which the instrument is operated. The instrument enters this mode when
the power is turned ON.

• Setting Mode

This mode is for entering the input range, alarms, and other settings.
• Hold down the

MENU

key for three seconds or longer while in operation mode to

enter this mode.

• To return to operation mode, hold down the

MENU

key for three seconds again.

• In this mode, the functions of the keys are those printed above the keys in

characters and symbols.

RCD

MENU

FEED

CH UP

FUNC

MENU

DISP

1

CHARACTER

ESC/?

SHIFT

• Basic Setting Mode

This mode is for setting up the thermocouple burnout function, action of alarm output
relays, and other of the recorder’s basic specifications.

• To enter this basic setting mode, hold down the

DISP

key and the

FUNC

key for

three seconds or longer while in setting mode. In this mode, the functions of the
keys are those printed above the keys in characters and symbols.

• To exit basic setting mode, display Basic = End then press the

key. Choose

Store or Abort in the End = screen, and press the

key. If you choose Store,

the settings are applied and the instrument returns to operation mode. If you
choose Abort, the settings are canceled and the instrument returns to operation
mode.

Note

For operating procedures in setting mode and basic setting mode, see the µR10000 Recorder
User’s Manual (IM04P01B01-01E).

• Test/Adjustment Mode

The mode in which the instrument is adjusted and tested.

• Enter this mode by pressing the

MENU

and

CH UP

keys while turning ON the power.

The Test/Adjustment Mode menu is displayed. In this mode, the functions of the
keys are those printed above the keys in characters and symbols.

• To return to operation mode, display Menu = End then press the

key.

• The following shows the structure of Test/Adjustment Mode. The shaded area is

not covered by this service manual.

CAUTION

Do not enter settings in Test/Adjustment Mode that are not specified by this
manual. Changing the settings can result in incorrect operation of the
instrument.

3.3 Performance Tests

This manual is related to the following products: