Keys, Power supply frequency detection, Battery backup – Yokogawa µR20000 User Manual

Page 26: Keys -13

background image

3-13

SM 04P01B01-01E

Testing

3

Keys

Procedures

1.

Turn ON the power switch while pressing the

MENU

and

CH UP

keys. Menu = A/D

Adjust is displayed.

2.

Select Menu = Test, and press the

key.

3.

Choose Test = Key and press the

key.

Confirm that the appropriate box turns On or OFF when each key is pressed.
The box of the display corresponds to the order of the keys.
Press the

ESC/?

key twice to return to the Test = Key screen.

Enter 2 ESC, Out.

* Select using the key. You can hold down the SHIFT key while pressing key to reverse the
key’s action.

Power Supply Frequency Detection

Procedures

1.

Turn ON the power switch while pressing the

MENU

and

CH UP

keys. Menu = A/D

Adjust is displayed.

2.

Select Menu = Test, and press the

key.

3.

Choose Test = Frequency and press the

key.

The detected power supply frequency is displayed. Confirm that it matches the
actual power supply frequency.

Frequency
50.0Hz

* Select using the key. You can hold down the SHIFT key while pressing key to reverse the
key’s action.

Battery Backup

Procedures

1.

Set the date and time on the recorder. Also, set an arbitrary channel to the 20
mV range.

2.

Turn OFF the power switch.

3.

Turn ON the power after approximately one minute, then check that the set
date, time, and range are not initialized.