beautypg.com

Channel-to-channel error, Burnout detection, Channel-to-channel error -11 – Yokogawa µR20000 User Manual

Page 24: Burnout detection -11

background image

3-11

SM 04P01B01-01E

Testing

3

Channel-to-Channel Error

Procedures

Pen Model
Assign inputs as shown in the table below in parallel to channels 1 and 2, and check
whether the measured values on both channels meets the criteria. Perform the same
test using channels 3 and 4.
Criteria

Range

Input

Criteria

20 mV

Short +/A and −/B

0.00 ±0.01 mV

Pt100

103.9 Ω

10 ±0.2°C, between channel dispersion within 0.2°C

Dot Model
Assign inputs as shown in the above table in parallel to channels 1 through 6, and
check whether the measured values on all channels meets the test criteria.

*

The b terminal of the resistance temperature detector input is connected internally.
However, when option /N2 is specified, the b terminal is isolated at each channel.

Burnout Detection

Procedures

The test is performed on a representative channel

*

from each A/D converter. Set up

the burnout detection function on the recorder in advance.

* On a four-pen model for example, perform the test on channel 1 and channel 3. On a dot

model, check an arbitrary channel.

Check the action in the following cases.
• Connect 2 kΩ between the channel’s +/A terminal and −/B terminal. Set that

channel’s thermocouple type to K and check that a value close to room
temperature is displayed.

• Connect 82 kΩ and 0.1 µF in parallel between the channel’s +/A terminal and −/B

terminals. Set that channel’s thermocouple type to K and check that the displayed
value is “B. out.”

3.4 Test Procedures

This manual is related to the following products: