Yokogawa EXA DC402 Dual-Channel Conductivity Analyzer User Manual

Page 71

background image

IM 12D08E02-01E

10-8 Appendix

Setting up DC402G for USP

First enable USP in service code 57. Change

the setting from 0 (default) to 1 (enabled).

This activates uncompensated conductivity in

the display menu. The E13 alarm feature is

also enabled. For E13 the FAIL alarm (S4) is

triggered when the uncompensated conductivity

exceeds the relevant value in the graph.

USP warning alarms can be selected for any of

the 4 relay outputs.

Service codes 40 to 43 are for these alarms.

Set to 2.5.1 for input one, or 2.5.2 for input two.

The setpoint is the safety margin in %, and is

set as described in sections 5-1-3 & 5-2-2.

Conductivity limit as a

function of Temperature

Temperature in ºC

0 25 50 75 100

μS/cm

3.5

3

2.5

2

1.5

1

0.5

0

Fig. 10-1.