3. calibration with hold active, 3. calibration with hold active -3 – Yokogawa EXA DC402 Dual-Channel Conductivity Analyzer User Manual
Page 56

IM 12D08E02-01E
Calibration 6-3
6-3. Calibration with HOLD active
mS / c m
HOLD
mS / c m
ENT
HOLD
mS / c m
ENT
HOLD
>
Select the flashing digit with the
> key.
Increase its value by pressing the
key
>
Set the value
using the
>, , ENT key.
mS / c m
NO
YES
HOLD
YES
mS / c m
NO
YES
HOLD
Put the sensor in standard
solution. Press
YES.
When the correct value is displayed,
press
ENT to enter the change.
After briefing displaying WAIT,
the CAL.END message appears.
The calibration is now complete. Put the
sensor back in the process and press
YES.
mS / c m
NO
YES
HOLD
HOLD will be displayed. Press
NO to turn off
HOLD and return to the measuring mode.
mS / c m
NO
YES
HOLD
mS / c m
NO
YES
NO
HOLD
NO
MODE
YES
ENT
YOKOGAWA
MODE
CONTACTS
S1
S2
S3
FAIL/S4
mS / c m
Press the
MODE key.
The legend CAL.1/CAL.2
appears, and the
YES/NO
key prompt flags flash.
If the SETP legend
appears, press
NO first.
MODE
MEASURE
DISPLAY 1
HOLD
DISPLAY 2
CAL 2
CAL 1
Note: See section 4-5 for full
display access function