3. calibration with hold active, 3. calibration with hold active -3 – Yokogawa EXA DC402 Dual-Channel Conductivity Analyzer User Manual

Page 56

background image

IM 12D08E02-01E

Calibration 6-3

6-3. Calibration with HOLD active

mS / c m

HOLD

mS / c m

ENT

HOLD

mS / c m

ENT

HOLD

>

Select the flashing digit with the

> key.

Increase its value by pressing the

key

>

Set the value

using the

>, , ENT key.

mS / c m

NO

YES

HOLD

YES

mS / c m

NO

YES

HOLD

Put the sensor in standard

solution. Press

YES.

When the correct value is displayed,

press

ENT to enter the change.

After briefing displaying WAIT,

the CAL.END message appears.

The calibration is now complete. Put the

sensor back in the process and press

YES.

mS / c m

NO

YES

HOLD

HOLD will be displayed. Press

NO to turn off

HOLD and return to the measuring mode.

mS / c m

NO

YES

HOLD

mS / c m

NO

YES

NO

HOLD

NO

MODE

YES

ENT

YOKOGAWA

MODE

CONTACTS

S1

S2

S3

FAIL/S4

mS / c m

Press the

MODE key.

The legend CAL.1/CAL.2

appears, and the

YES/NO

key prompt flags flash.

If the SETP legend

appears, press

NO first.

MODE

MEASURE

DISPLAY 1

HOLD

DISPLAY 2

CAL 2

CAL 1

Note: See section 4-5 for full

display access function