7. configuration checklist for dc402g, 7. configuration checklist for dc402g -6 – Yokogawa EXA DC402 Dual-Channel Conductivity Analyzer User Manual
Page 69

IM 12D08E02-01E
10-6 Appendix
10-7. Configuration checklist for DC402G
Standard Configuration
Options
Reference for
change
Measured Variable(s)
primary inputs
Conductivity (SC) and Temp Resistivity i.s.o Conductivity
code 01
conductivity range
0.000 µS/cm - 100 µS/cm
any span within 0.000µS/cm - 1999mS/cm
“range”
conductivity units
Auto ranging µS/cm - mS/cm Choice out of fixed µS/cm or mS/cm
code 56
resistivity range
0 - 19.99 MΩ•cm
any span within 0.000kΩ•cm - 999MΩ•cm
“range”
resistivity units
MΩ•cm
Auto ranging or other fixed values
code 56
temperature range
0 - 100 °C
any span in -20 ... +250
“range”
temperature unit
Celsius
Fahrenheit
code 11
Outputs
analog output
4- 20 mA for SC
0- 20 mA or 4- 20 mA
code 30
second output
4- 20 mA for Temp
0- 20 mA or 4- 20 mA
code 30
output allocation
SC and Temp
SC, Resistivity, Temp, Table, PI control
code 31
contact outputs
S1= high at 100 µS/cm
(4) freely programmable
“setpoint”
S2= low at 100 µS/cm
code 40. 41, 42, 43
S4= FAIL
contact allocation
mS/cm and FAIL
µS/cm, mS/cm, kΩ•cm, MΩ•cm,
code 40- 43
temp, PI control, HOLD, FAIL
contact variables
dead time= 0.2 s; hyst= 0.1% time: 0- 200 s; hyst 0.1- 100%
code 44
add. contact functions
none
time out alarm
code 47
control functions
none
PI on contacts or mA output
code 45, 46, 34, 33
digital outputs
none
RS485
code 60
Communication
digital interface
disabled
RS485
code 60
variables on display
µS/cm (mS/cm) and temp
TC, %w/w, °C, mA1, mA2, CC, TR, REL
“display”
burn out
disabled
burn low (3.5)/ high (22) on mA1/ mA2
code 32
password protection
disabled
for maint/ comm./ serv level
code 52
autoreturn
return to measure in 10 min. enable or disable
code 50
add. function in MAINT
disabled
setpoint adj
code 51
Diagnostics
check on polarization
enabled
enable or disable
code 05
check on cell constant
active
“calibrate”
check on TC coefficient
active
“temp.”
Compatibility
SC sensor
SC40, SX42
SC8SG, L&N compatible
code 10
temperature sensor
Pt1000
Ni100, PB36, Pt100, 8k55
code 10
cell constant
0.100/cm
from 0.008/cm up to 50.0/cm
code 03
Special Features
temperature calibration
none
adjustment +/- 15 °C
code 12
zero calibration
none
adjustment -1 µS/cm
code 04
temp. comp.
acc to NaCl tables IEC 60746-3 NaCl, manual TC, matrix
“temp.”,
code 20 - 28
HOLD during maintenance
disabled
hold last or hold fix
“Hold”
contact during HOLD
disabled
possible on S1,S2 or S3
code 40- 42
soft fail alarm
disabled
possible for E1, E5...E13, E22
code 53