Calibration, 1. calibration check with buffer solutions, 2. manual calibration mode – Yokogawa EXAxt PH450 4-Wire Analyzer for pH and ORP User Manual
Page 48: 3. automatic calibration mode

40
IM 12B07C05-01E
6-1. Calibration check with buffer solutions.
The following tips will help to produce a good
calibration.
1. Before starting a calibration, make sure the
electrode system is properly cleaned so that
the electrodes are fully functional. They must
then be rinsed with clean water to avoid con-
tamination of the calibration solution(s).
2. Always use fresh buffer solutions to avoid the
possibility of introducing errors from contami-
nated or old solutions. Buffers supplied as liq-
uids have a limited shelf life, especially alkaline
buffers, which absorb CO
2
from the air.
3. Yokogawa strongly recommends NIST
(primary) buffer standards in order to ensure
the best accuracy and best buffer capacity is
available. Commercially adjusted buffers (e.g.
7.00, 9.00 or 10.00 pH) are a compromise as
a standard, and are often supplied without the
temperature dependency curve. Their stability
will never be as good as NIST solutions.
Always ensure that the sensors are properly
conditioned, clean and filled with the correct
electrolyte solution (if appropriate) before starting
a calibration. Refer to section 7 (Maintenance),
and to the sensor instructions for details.
6-2. Manual calibration mode
The unit is adjusted to agree with the value of
a known solution. This may be a buffer solution
or a known process sample. The user deter-
mines the pH value, the temperature influence
and the stability.
1- A single point can be set to adjust only the
zero (asymmetry).
2- A second point can be set to determine the
slope (sensitivity).
6-3. Automatic calibration mode
The PH450G will provide prompts to aid the
user to make a good calibration. High quality
buffer solutions must be used for best results.
The user selects the buffer type that he is
using in the calibration menu. The buffer set is
selected in
Commissioning >> Measurement
setup >> Calibration setting>> Buffers
See also Appendix 1.
The PH450G uses temperature sensor input to
determine the exact buffer values.
The EXAxt also determines the stability (drift)
and will reject the new calibration data if it is
outside limits.
The PH450G records the values internally, and
uses them to calculate the final calibration.
1- A single point can be set to adjust only the
zero (asymmetry).
2- A second point can be set to determine the
slope (sensitivity).
6. CALIBRATION