beautypg.com

3-5. grounding the housing, 3-6. switching on the instrument, Warning – Yokogawa EXAxt PH450 4-Wire Analyzer for pH and ORP User Manual

Page 18: Danger, Power

background image

10

IM 12B07C05-01E

3-3-5. Grounding the housing

For the safety of the user and to protect the

instrument against interference, the housing

must always be connected to ground. This has

to be done by a large area conductor. This

cable can be fixed to the rear of the housing or

by using the internal ground connections using

a braided wire cable. See figure 3-8.

The minimum cross sectional area of the pro-

tective grounding wire should be 0.75 mm

2

.

3-3-6. Switching on the instrument

After all connections are made and checked,

the power can be switched on from the power

supply. Make sure the LCD display comes on.

After a brief interval, the display will change to

the measured value. If errors are displayed or

a valid measured value is not shown, consult

the troubleshooting section (Chapter 8) before

calling Yokogawa.

DANGER

Figure 3-7. Input and output connections

2

1

-

+

DC

POWER

12-24 V /10 W

FUSE: 2A/250 VAC/T

2

1

N

L

AC

POWER

100-240 VAC/15 VA/ 50/60Hz

FUSE: 500 mA/250 VAC/T

Fuse replacement should be performed only by

a qualified service personnel.

See Sec.7. MAINTENANCE, Fuse.

Fuse ratings:

Power supply

Fuse type

12-24 VDC, 10W max

2A/250V, Slow

100-240 VAC, 15VA max 0.5A/250V, Slow

3-3-2. Access to terminal and cable entry

Terminals 1 and 2 are used for the power sup-

ply. Guide the power cables through the gland

closed to the power supply terminals. The ter-

minals will accept wires of 2.5 mm

2

(14 AWG).

Always use cable finishings if possible.

WARNING

3-3-3. AC power

Connect terminal L to the phase line of the AC

power and terminal N to the zero line. See fig-

ure 3-8 for the power ground. This is separated

from input ground by a galvanic isolation.

3-3-4. DC power

Connect terminal 1 to the positive outlet and

terminal 2 to the negative outlet.

This is separated from input ground by a

galvanic isolation. The size of conductors

should be at least 1.25 mm

2

. The overall cable

diameter should be between 6 & 12 mm.