4-7. Navigation of the menu structure

“HOME KEY” return to main display

“RETURN

KEY”

exit

to

previous

display

Main display

Primary setup display

Commisioning menu

Instrument in HOLD

4-8. Setup Concentration mode

The concentration mode allows the user to

generate an analog output signal that is linear to

concentration units and to read the concentra-

tion on the LCD in the units %, mg/l or ppt.

Example: PH450 is used as ORP analyser and

the output is linear to mg/l Free Chlorine.

As first step the table for mA1 must be filled

in.

Then the concentration menu is opened: and

begin and end value of the scale are entered

Setup is completed.

The measured concentration is now displayed

on the top line of the LCD.