Caution – Yokogawa EXAxt PH450 4-Wire Analyzer for pH and ORP User Manual
Page 12

4
IM 12B07C05-01E
CAUTION
This instrument is a Class A product, and it is designed for use in the industrial environment.
Please use this instrument in the industrial environment only.
O) Environment and operational conditions
Ambient temperature : -20 to +55 ºC (-5 - 130 ºF)
Storage temperature : -30 to +70 ºC (-20 - 160 ºF)
Humidity
: 10 to 90% at 40 ºC (100 ºF) RH (non-condensing)
Data protection
: EEPROM for configuration data and logbook. Lithium cell for clock
Watchdog timer
: Checks microprocessor
Power down
: Reset to measurement
Automatic safeguard: Auto return to measuring mode when touchscreen is untouched for 10 min
Model and Suffix Codes
[Style: S2]
Model
Suffix code
Option code
Description
PH450G
-----------------
-----------------
pH / ORP Converter
Power
- A
- D
-----------------
-----------------
AC version (100…240 VAC)
DC version (12…24 VDC)
Type
- A
- U
-----------------
-----------------
General purpose version
FM version
Mounting Hardware
Hood
Conduit adapter
Tag Plate
/UM
/U
/PM
/H5
/AFTG
/ANSI
/AM20
/SCT
Universal mounting kit (panel, pipe, wall)
Pipe and wall mounting hardware (*2)
Panel mounting hardware
(*2)
Awning hood (stainless steel) (*2)
G1/2 (*2)
1/2NPT (*2)
M20 (*2)
Stainless steel tag plate (*1)
Notes:
*1 If the tag number is predefined with the purchase, Yokogawa will inscript the tag plate with the specified tag number, and
program the tag number in the converter.