beautypg.com

Operation of exaxt ph450g, 1. main display functions, 2. trending graphics – Yokogawa EXAxt PH450 4-Wire Analyzer for pH and ORP User Manual

Page 27: 3. zoom in on details, 3-1. actual ma1, 3-2. actual ma2, 3-3. zero, Next

background image

19

IM 12B07C05-01E

4

OPERATION

OF
EXAxt

PH450G

4-1. Main display functions

T ag:EXAxt PH45 0

HOLD

W AS H

25.0

236.0 mV

S4

S3

S2

S1

Go to trend screen

Go to zoom screen
Go to status screen

Go to maintenance screen

Figure 4-1. Main Display

A heart “ ” mark is lit on the right-upper corner of the
screen when HART communication is active. An “Χ”
mark is lit when HART communication is abnormal.
Nothing appears when HART communication is not
used.

Note that the “Χ” mark may appear due to output

signal noise or the like even when HART communica-

tion is not used. There is not problem when HART

communication is not used. Continue operation while
ignoring the mark.

4-2. Trending graphics

Pressing the button changes the display

into a graphical mode in which the average

measured value is shown on a time scale. The

“Live” value is also digitally displayed in a text

box. The time scale ( X-axis) and the primary

value scale (Y-axis) are set in the “DISPLAY

SETUP” menu. The full screen displays a trend

of 51 points that represent the average of the

selected time interval. The converter samples

the measurement every second. The trending

graphic also shows the maximum and minimum

measured value in that interval.

For example if the time scale is set to 4 hours,

then the trend is shown for 4 hours prior to the

actual measurement. Each point on the trend

line represents the average over 4*60*60/51 =

282 measurements (seconds).

Maximum

Minimum

Minimum

Maximum

Average

Live reading

9.10 pH

4

Figure 4-2. Trend screen

4-3. Zoom in on details

This button gives access to the diagnostic

information of the converter. The following

messages will appear under normal (default)

conditions:

- Home key back to
mainscreen.
- One level up.

- Scroll choices
(grey

means deactivated).

- Enter selected data
or choice.

First zoom screen
gives you inside
into the parameters
involving current
measurement. All
following zoom
screens give
additional information
about the device and
lead to logbook data.

Next

Next

Next

Next

Figure 4-3. Detail screen

4-3-1. Actual mA1

= the current output in mA of the first current

output, which is defined as mA1. The range

and function of this mA output can be set in

Commissioning >> Output setup>> mA1

4-3-2. Actual mA2

= the current output in mA of the second

current output, which is defined as mA2. The

range and function of this mA output can be

set in

Commissioning >> output setup >> mA2

4-3-3. Zero

= calibrated sensor offset in mV. Theoretically

the sensor reads 0 mV in a buffer solution with

the same pH value as the Iso po ten tial pH value

4. OPERATION OF EXAxt PH450G

T

-->

pH

--

>