Vanguard CT-6500 S2 User Manual
Page 58

CT-6500 Series 2 Operating Instructions
57
9.14 Transducer Self-Test
The operator can check the transducer electronics by connecting the transducer to the CT-
6500 (D1 or D2 or D3) and running a “Test Transducer” diagnostics.
Table 19.0 Performing a Transducer Self Test
STEP OPERATION
ACTION
DISPLAY
1 SELECT
DIAGNOSTIC.
PRESS # 4 KEY.
1. SLOW-CLOSE TEST
2. CHECK HOOKUP
3. TEST TRANSDUCER
4. NEXT PAGE
2
SELECT TRANSDUCER.
PRESS # 3 KEY.
TRANSDUCER TEST
1= 0.00 IN / 000.0
°
2= 0.00 IN / 000.0
°
3= 0.00 IN / 000.0
°
3
MOVE TRANSDUCER SLIDER TO A
KNOWN DISTANCE & OBSERVE
DISPLAY
MOVE
TRANSDUCER
SLIDER
TRANSDUCER TEST
1= 5.00 IN / 360.0
°
2= 0.00 IN / 000.0
°
3= 0.00 IN / 000.0
°
2
RETURN O MAIN MENU
PRESS # STOP
KEY.
1.RUN TEST 06/10/07
2.GET RESULTS 20:10:01
3.SETUP
4.DIAGNOSTIC
NOTES
1. After selecting the transducer test mode, move the transducer slider and observe the
display on the CT-6500 LCD.
2. If the slider is moved 1.00 inch, the CT-6500 LCD should display 1.00 inch.
3. Press the “STOP” key to return back to the main menu.