Vanguard CT-6500 S2 User Manual
Page 45

CT-6500 Series 2 Operating Instructions
44
9.6 Recalling a Timing Test Record
The following steps in Table 11.0 are used to recall a timing test record stored in the EEPROM
to the working memory. Once a test record is recalled, the user can get a printout of the test
results from the thermal printer.
Table 11.0 Recalling a Timing Test Record
STEP OPERATION
ACTION
DISPLAY
1
SELECT SETUP MENU.
PRESS #3 KEY.
1.ANALYSIS POINT
2.MEASUREMENT UNITS
3.SAVE / RESTORE
4.NEXT PAGE
2
SELECT SAVE/RESTORE SHOT.
PRESS # 3 KEY.
1.SAVE SHOT
2.RESTORE SHOT
3.SHOT DIRECTORY
4.ERASE SHOT
3
SELECT RESTORE SHOT.
PRESS #2 KEY.
ENTER SHOT NUMBER
TO BE RESTORED
XXXX
"ENTER" TO CONFIRM
4
ENTER SHOT NUMBER “1” TO BE
RESTORED.
PRESS KEYS 0-9
FOR SHOT ID THEN
PRESS THE
"ENTER" KEY.
ENTER SHOT NUMBER
TO BE RESTORED
0001
"ENTER" TO CONFIRM
5
CONFIRM SHOT TO BE RESTORED.
PRESS THE
"ENTER" KEY.
RESTORING SHOT # 0001
TEST: OPEN
DATE: 06\12\07 09:12
"ENTER" TO CONFIRM
6
CONFIRM TEST RESTORED.
NONE.
RESTORE COMPLETE
7
RETURN TO MAIN MENU.
PRESS ANY KEY.
1.RUN TEST 06/10/07
2.GET RESULTS 20:10:01
3.SETUP
4.DIAGNOSTIC