beautypg.com

Probe calibration with the tf-dsq fixture, Probe calibration steps with the tf-dsq fixture, Tf-dsq overview – Teledyne LeCroy TF-DSQ User Manual

Page 6: Operator's manual

background image

Operator's Manual

6

TF-DSQ-OM-E RevC

TF-DSQ Overview

Probe Calibration with the TF-DSQ Fixture

The TF-DSQ fixture is used in conjunction with the oscilloscope software to perform probe
deskew and DC calibration. It has the following leading features and specifications:

Deskew to +/- 20 ps typical accuracy

Differential and single-ended drive

75 ps edge

Calibration of DC gain and offset and skew at same probing point

Accounts for risetime variations

Accounts for common-mode voltage

DC gain calibration accounts for probe loading effects

Integrated operation with oscilloscope for fully automatic calibration

WavePro 7 Zi and

WaveMaster 8 Zi

Oscilloscopes and SDA/DDA

7 and 8 Zi Models

WavePro 7000

Oscilloscopes

WaveMaster 8000

Oscilloscopes and SDA/DDA

Models, also DDA 5005A,

5005A XXL

Oscilloscope

Connection

ProLink and ProBus

ProBus Only

ProLink Only

Probes

Supported

D610,D620, D600A-AT, D310, D320, and D300A-AT with all probing accessories

AP020, AP033, AP034 HFP1000, 1500, 2500, and 3500, PP005 and PP005A

DC Range

±5 V single-ended, ±10 V differential

DC Accuracy

±(1% + 600 μV)

Edge Risetime

(of Device)

75 ps (typical) < 95 ps (guaranteed)

Edge Amplitude

and Rep Rate

Approximately 800 mV @ 10 MHz

Deskew

Accuracy

±20 ps (typical)