Deskew only, The advanced menu, Rise time skew correction – Teledyne LeCroy TF-DSQ User Manual
Page 18: Deskew all (or common skew), Operator's manual
![background image](/manuals/353494/18/background.png)
Operator's Manual
18
TF-DSQ-OM-E RevC
Deskew Only
Pressing this button performs only the deskew calibration of the probe on the specified channel.
See details
The Advanced Menu
The Advanced Menu contains information and functionality useful to the advanced user of the
TF-DSQ fixture. These include:
Common mode voltage settings for DC calibration
Rise Time Skew Correction
This field shows the signal risetime and the corresponding skew correction based on the signal
risetime.
When probes are deskewed, the risetime measurement of the edge used for deskewing is
displayed in the Rise Time field corresponding to the probe and probe channel, and an
additional skew correction of zero is applied.
The measured risetime of the signals encountered can be entered into the Rise Time field, and
the oscilloscope automatically calculates and applies a new skew correction value to be utilized
in addition to the deskew amount calculated during the deskew calibration procedure. With this
use, a finer deskew calibration is performed because the risetimes of the signals measured are
accounted for. See details of
Deskew All (or Common Skew)
This is the deskew amount applied to all channels. The time entered in this field is the absolute
time by which all waveforms displayed by the oscilloscope are delayed in time. This value
effectively adjusts the zero time reference of the system. See
details.