Specifications

PXD222-OM-E Rev A

ISSUED: November 2001

31

SAFETY

The LeCroy PXD 222 Digitizer is designed for measurements on 300 V Category II Installations,
Pollution Degree 2, per:

EN 61010-1:1993 + A2:1995

CAN/CSA-C22.2 No.1010.1-92

UL3111-1

Max. Input Voltages:

Input Ch 1 and Ch 2 directly.................... 300 V CAT II

Input Ch 1 and Ch 2 via 10:1 probe ........1000 V CAT II

Max. Floating Voltage:

From any terminal to ground................... 300 V CAT II

Between any terminal.............................. 300 V CAT II

Voltage ratings are given as “working voltage.” They should be read as VAC rms (50–60 Hz) for
AC sine wave applications and as VDC for DC applications.

CAUTION

Safety specifications are only valid if
the unit is properly grounded, as
described in the Installation procedure.