Specifications
PXD222-OM-E Rev A
ISSUED: November 2001
29
Trigger Sensitivity Ch1 and Ch2:
DC to 5 MHz............................... 5 mV or 6.25% of full range
@ 200 MHz ............................... 12.5% of full range
@ 250 MHz ............................... 25.0% of full range
Video Trigger:
Standards...................................PAL, PAL+, NTSC, SECAM
Modes ........................................ Lines, Line Select, Field 1, or Field 2
Source........................................ Ch 2
Polarity .......................................Positive, Negative
Sensitivity (sync level) ................ 8.75% of full range
Pulse Width Trigger:
Screen Update ........................... On Trigger, Single Shot
Trigger Conditions ......................< T, > T, approximately equal to T (±10%), not equal to T
(±10%)
Source........................................ Ch2
Polarity .......................................Positive or negative pulse
Pulse Time Adj. Range .............. 0.25 to 10 samples with a maximum resolution of 50 ns
Auto Set/ Range:
Autoranging attenuators and timebase; automatic triggering with automatic source selection. The
autorange must be switched on or off.
If autorange is on, all auto functions are active until an autorange off command is executed.
Modes:
Normal........................................ 15 Hz to max. bandwidth
Low Frequency........................... 1 Hz to max. bandwidth
Minimum Amplitude Ch 1, Ch 2
DC to 1 MHz............................... 20 mV
1 MHz to max. bandwidth........... 40 mV