beautypg.com

Deskewing master and slave channels – Teledyne LeCroy Zi Oscilloscope Synchronization ProBus Module (Zi-8CH-SYNCH) User Manual

Page 9

background image

Instructions 

 

918246‐00‐RevA 

9

Deskewing Master and Slave Channels

Deskew Theory of Operation

Deskewing is an adjustment of waveform data point occurrences on the screen. Deskewing is an operation to 
correct the times waveforms are displayed on the screen, mainly to account for propagation delays through 
probes and cables, and transfer times from the Slave to the Master oscilloscope. When considering skew, there 
are two important things to consider: 

• The relative skew between two channels. 
• The absolute skew from the zero time reference (the trigger point). 

Two channels are properly deskewed relative to each other when the difference between the deskew values 
entered for each channel aligns an edge occurring at the same time and applied to both channels. A channel is 
properly deskewed in an absolute sense when an edge is applied to the channel, the oscilloscope is triggered on 
the edge, and the edge appears such that the trigger threshold crossing appears at the trigger delay, which is the 
zero time reference on the oscilloscope screen. 

Probes or cables are deskewed one at a time relative to a reference such that the resulting calibration deskews 
each probe relative to every other probe or cable. 

The WaveMaster 8 Zi has a built‐in fast edge SMA output. This edge may be used as a common deskew 
reference edge, and output to either the SMA power splitter (provided) and then to cables connected to the 
oscilloscope channels, or to the PCF‐200 Probe Deskew Fixture provided with many LeCroy probes (if probes are 
used in your measurements). Or, users can provide a fast edge signal of their own. Regardless, the exact cable or 
probe setup, and exact V/div range used in actual signal measurements should be used in the deskew procedure 
to ensure the most accurate and repeatable results. 

Note: The exact cable setup you plan on using for actual measurements from step 6 of Signal Connections (on 
page 4) must be the same setup used along with the cable splitter for the following deskew procedure for 
Master and Slave oscilloscope channels. 

During a deskew procedure when using the Zi‐8CH‐SYNCH, the Master and Slave oscilloscopes must first be 
deskewed with respect to each other using a common reference edge. This establishes a channel + probe/cable 
reference point within both the Master and the Slave oscilloscopes. This same reference point is then used to 
deskew all channels within each oscilloscope. Typically, the deskew values required are very small, but it is good 
engineering practice to verify proper deskew and adjust as needed. 

PLEASE NOTE THE FOLLOWING:  

• If you're using probes on your input signals, be sure to include them in your deskew measurements. 
• When providing your reference signal for deskewing signals, make sure it matches the appropriate V/div 

range you plan on using for your actual measurements. 

Creating Master and Slave Deskew Channel References

Use the following procedure to first establish a deskew signal reference between C2 on the Master and C2 (or 
the relabeled C6)
 on the Slave oscilloscope. 

1. Set up the Zi‐8CH‐SYNCH to trigger in Periodic mode. 

Note: In Periodic mode you can only deskew by acquiring A SINGLE ACQUISITION at a time.