Operating Environment

Before using your LeCroy accessory, ensure its operating environment will be maintained within these 
parameters: 

• Indoor use; 
• Altitude up to 3000 m; 
• Temperature 5 °C to 40 °C; 
• Maximum relative humidity 80 % for temperatures up to 31 °C decreasing linearly to 50 % relative 

humidity at 40 °C. 

General Safety Summary

Review the following safety precautions to avoid injury and prevent damage to your LeCroy accessory or any 
products connected to it. 

Use only as intended. This accessory is intended to be used only with the compatible LeCroy instruments. 

Use of this product and/or the equipment it is connected to in a manner other than specified may impair 
the protection mechanisms. 

Connect and disconnect properly. Avoid damage thru excessive bending. 
Do not use in wet/damp or explosive atmospheres
Do not operate with suspected failures. Do not use this accessory if any part is damaged. All 

maintenance should be referred to qualified service personnel. 

Keep product surfaces clean and dry

Compliance Information

This LeCroy accessory is CE Compliant and bears CE mark. 

EC Declaration of Conformity

The product meets intent of the European Council Directive 2006/95/EC for Product Safety and 2004/108/EEC 
for Electromagnetic Compatibility. Compliance was demonstrated to the following specifications as listed in the 
Official Journal of the European Communities: 

• EN 61010‐1:2001 Safety requirements for electrical equipment for measurement, control and laboratory 

use. 

• EN 61326‐1:2006 EMC requirements for electrical equipment for measurement, control and laboratory 

use. 

• EN 55011/A2:2002 Radiated and Conducted Emissions, Group 1, Class A 

1

 

• EN 61000‐4‐2:2001 Electrostatic Discharge Immunity (±4 kV contact discharge, ±4 kV air discharge, 

±4 kV vertical/horizontal coupling planes) 

• EN 61000‐4‐3:2006 RF Radiated Electromagnetic Field Immunity (3 V/m, 80 MHz to 1 GHz; 3 V/m, 

1.4 GHz to 2 GHz; 1 V/m, 2 GHz to 2.7 GHz , 80% amplitude modulated with 1 kHz sinewave) 

 

1

 This product is intended for use in nonresidential areas only. Use in residential areas may cause 

electromagnetic interference.