Instructions 

 

918246‐00‐RevA 

13

Now, align the edges of your acquired signals by fine adjusting the deskew value of the C2 (Master) channel (the 
one previously deskewed). 

Note: Do not adjust the deskew of the C2 (relabeled C6, Slave) Channel since it has a zero value and is the 
reference for all other adjustments. Depending on your application, it may also be possible to perform this 
single‐shot fine deskew adjustment using a customized routine (refer to the Customization sections covering 
XDEV in the online help on your oscilloscope for more details). 

These last fine adjustments make it possible to deskew the Master and Slave with reference to each to within ~1 
ps
. Keep in mind that one single long acquisition using this additional fine deskew adjustment does not take any 
more time than shorter acquisitions, yet it returns an equivalent or better set of data in many measurement and 
analysis scenarios.