Instructions – Teledyne LeCroy Zi Oscilloscope Synchronization ProBus Module (Zi-8CH-SYNCH) User Manual

Page 11

background image

Instructions 

 

918246‐00‐RevA 

11

The highlighted section of the following figure illustrates this use of the module. 

 

Tip: Turning on WaveStream or persistence during deskew adjustment provides an ideal display 
for visual inspection of the deskew adjustment accuracy. 

• In User Edge trigger mode, it’s easier to measure deskew accuracy with the Skew parameter 

(shown as follows) over multiple acquisitions. From the menu bar, select Measure → Measure 
Setup
. On the P1 dialog, select your inputs C2 and C6 (C2 from the Slave), respectively, and the 
Skew measurement parameter from the pop‐up window. 

 

At this point, go to the main Measure dialog and mark both the Show Table and Statistics 
checkboxes. You can then easily view minmax, and sdev measurements to ensure your deskew 
adjustments are within an acceptable range. 

The Skew parameter should return an sdev value of ~2.5ps and the min and max values should be 
equally distributed around zero. For most accurate results, use as fast an edge as possible.