Section 4: Service & Maintenance

04/10

31

51205 Rev. 2

5.3. SCALE PLATFORM TROUBLESHOOTING

Except for severe structural damages, most Platform
Assembly problems can be traced to the following causes.

• Material under or around the Platform
• Incorrect, loose or damaged Load Cells

5.3.1. Scale Platform Testing

5. Inspect the Interface Cable from the Platform to the Instrument for visible breaks

or cracks.

6.

ZERO

the Instrument Display.

7. Apply a test load of

25% of the Load Cell capacity

to one corner.

─ The Instrument should display a weight reading within 0.1% of the applied

weight, or One Instrument Division, whichever is greater.

8. Repeat Step 3 for all the corners, placing the same Test Load on each corner.

5.3.2. Load Cell Testing

When corners do not match the correct tolerances, disconnect the leads from the
summing card’s terminal strip and test the load cell using the settings on the following
chart.

.

TEST

READING

REMARKS

Green to Black (Input)

1106 Ohms (+5 / -2 Ohms)

Input Resistance

Red to White (Output)

1000 Ohms (+5 / -2 Ohms)

Output / Bridge Resistance

Yellow (Shield) to Load Cell Case

More than 1,000 megohms

Insulation Resistance

Input and Output Leads to Shield

Input and Output Leads to Case