Section 4: Electrical Installation

04/10

23

51205 Rev. 2

4.1.1. Moisture Protection

The

LCF-4020 Rocker Column Load Cells

give full moisture protection.

• The load cells are calibrated with the cable attached.
• The cable

MUST NOT be cut.

• The cable is connected directly to the Junction Box through a sealed bushing

which

MUST BE TIGHTENED WITH PLIERS

to keep moisture out of the box.

On all Junction Boxes, the black plastic fittings have
O-rings that can be forced out of position if the
bushing itself is not tight.

1. To prevent this, first tighten the inner nut securing

the bushing in the hole.

2. Then insert the cable and carefully tighten gland

with pliers until it is very snug.
DO NOT over-tighten where bushing 'turns'.

• All Junction Box covers

MUST BE SECURED

and

tightened.

4.1.2. Data Recording

Keeping records of work completed for every customer is extremely important, and
can pay off in great measure in future visits.

Record scale serial number from the tag.

Record instrument, junction box and load cell serial numbers.

Keep a copy of the sheet in the customer file.

An example of a “drip loop”.
This helps prevent water entry.