Total Organic Carbon Analyzer

Maintenance

Teledyne Analytical Instruments

55

5.3.5 CE Computer (P/N 13039)

If malfunctioning, replace the Computer Module using the

directions below. See Figure 5-7.

Figure 5-7: Replacing CE Computer

To Replace the CE Computer:

1. Turn off power
2. Disconnect electrical connectors.
3. Remove screws holding Module in place.
4. Remove module.
5. Install new CE Computer in reverse disassembly

instruction procedure steps. (Install electrical connectors
per supplied drawing).

6. Verify proper software install per directions shipped with

new CE Computer.

7. Recalibrate liquid zero and span values.